home

ABOUT JSA

expertise

projects/
portfolio

career
opportunities

contact jsa

         

Bio-Level IV Decontamination Tanks

Scope

Renovate the two existing 1,000-gallon Bio-Level IV decontamination tanks from a steam-jacketed system to a direct inject steam system.

decontamination tank

Owner 
Centers for Disease Control

Mechanical Contractor
Peachtree Mechanical

Prime Consultant
A&E Design

Construction Cost
$440,000


Special Circumstances

  • Phasing of the project to maintain operability of the BL-IV laboratories during the construction period.

  • Implementing load cells to determine real time tank capacities.

  • Structurally isolating the two tanks via flex connectors and piping expansion loops to ensure accuracy of the load cell systems.

  • Installation of suction tees to inject steam and agitate the waste products within the tank.

  • Design of a spore test tube to verify proper operation of the system.

  • Adhering thermocouples to the tank body to acquire an array of temperature readings for control sequencing.

  • Automating the controls system to include tank filling, cooking, venting, cooling, and draining.

   Copyright © 2010 Johnson, Spellman & Associates, Inc.